Calibration Procedure
Enter Calibration Mode
To begin the calibration process, log in using the Admin password. The factory default is no password.
Front Panel
SCPI
Select
System > Admin > Login
. Then enter the password and press
[Select]
.
CAL:STAT ON
[, <
password
>]
DC Voltage Low Range
Connect the voltage input of the 3458A DMM to the instrument output as shown below and configure
the 3458A for DCV measurements. You can use the following SCPI commands (or their front-panel
equivalents) to configure the DMM:
*RST
DCV AUTO, AZERO ON, FIXEDZ OFF
NPLC 100
Keysight AC6800B Series Operating and Service Guide
171
4 Calibration, Verification, and Service