Step
Description
Front Panel
SCPI
1
Select the low voltage
range.
[Menu] > System > Admin > Cal > DC
Then select the
Low
range.
CAL:VOLT:OFFS 219
2
Select the first voltage
calibration point.
Select
Next
. The information field should indicate:
Enter P1 measured data
CAL:LEV P1
*OPC?
3
Measure the output
voltage with the DMM
and enter the data.
Enter the measurement from the external DMM in
the
Measured Data
field. Press
[Select]
or
[Enter]
when done.
CAL:DATA <
data
>
4
Repeat steps 2 and 3 for calibration point 2 (P2). Note that the SCPI command in step 2 will change to
CAL:LEV P2, and the front-panel message will read
Enter P2 measured data
.
5
Select
Next
or press
Back
to finish this part of the calibration.
DC Voltage High Range
Use the same DMM configuration as for the DC Voltage Low Range calibration.
Step
Description
Front Panel
SCPI
1
Select the high voltage
range.
[Menu] > System > Admin > Cal > DC
Then select the
High
range.
CAL:VOLT:OFFs 438
2
Select the first voltage
calibration point.
Select
Next
. The information field should
indicate:
Enter P1 measured data
CAL:LEV P1
*OPC?
3
Measure the output
voltage with the DMM
and enter the data.
Enter the measurement from the external DMM
in the
Measured Data
field. Press
[Select]
or
[Enter]
when done.
CAL:DATA <
data
>
4
Repeat steps 2 and 3 for calibration point 2 (P2). Note that the SCPI command in step 2 will change to
CAL:LEV P2, and the front-panel message will read
Enter P2 measured data
.
5
Select
Next
or press
Back
to finish this part of the calibration.
AC Voltage Low Range
Use the same physical configuration that you used for DC Voltage calibration, but configure the 3458A
for AC voltage measurements. You can use the following SCPI commands (or their front-panel
equivalents) to configure the DMM:
*RST
SETACV ANA, RANGE AUTO, ACBAND 20, 2E6
ACV
4 Calibration, Verification, and Service
172
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...