DMM Configuration
For all tests besides the DC Voltage and Current tests, the DMM should be set to measure AC volts
and 100 NPLCs. For the DC voltage tests, the DMM should be set to measure DC volts and 100 NPLC.
Constant Voltage Tests
If more than one meter or a meter and an oscilloscope are used, connect each to the sense terminals
by separate leads to avoid mutual coupling effects.
4 Calibration, Verification, and Service
178
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...