Step Action
Normal Result
4
Program output to RANGE 310, VOLT 0, OUTP:COUP DC, OFFSET 128.
Output voltage at 128 VDC, CV annunciator
on, output current near:
2 A for AC6801B
4 A for AC6802B
8 A for AC6803B
16 A for AC6804B
5
Record DVM reading and calculate current.
Readings within specified dc current read-
back limits.
4 Calibration, Verification, and Service
184
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...