DISPlay Subsystem
The DISPlay subsystem controls the front-panel display.
DISPlay[:WINDow]:METer:COUPling AC|DC|ACDC
DISPlay[:WINDow]:METer:COUPling?
Selects the data to show on the meter display.
Parameter
Typical Return
AC|DC|ACDC
AC, DC, or ACDC
Set the meter to display DC measurements: DISP:MET:COUP DC
AC shows pure AC measurement data, DC shows DC measurement data, and ACDC shows all AC
and DC measurement data.
The
command sets this parameter to ACDC.
DISPlay[:WINDow][:STATe] ON|1|OFF|0
DISPlay[:WINDow][:STATe]?
Turns the front-panel display on or off.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Turn the front-panel display off: DISP:STAT OFF
Setting this OFF stops the refreshing of the display, turns off the display's backlight, changes the LINE
LED to orange, and disables all front-panel key operations.
The
command sets this parameter to ON.
DISPlay[:WINDow]:TEXT "<
string
>"
DISPlay[:WINDow]:TEXT?
Displays a text message on the display.
Parameter
Typical Return
Quoted string of up to 127 characters, default "". "Test running. Do not touch."
Show message on display: DISP:TEXT "Test in progress..."
While a message is displayed, instrument operation information is not sent to the front-panel display.
To clear the text, send the
command or cycle power.
The display text is unaffected by *RCL and *RST`.
Keysight AC6800B Series Operating and Service Guide
123
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...