INITiate Subsystem
The INITiate commands initialize the trigger system. This moves the trigger system from the "idle"
state to the "wait-for-trigger" state, which enables the instrument to receive triggers. An event on the
selected trigger source causes the trigger to occur.
INITiate[:IMMediate]:ACQuire
Initiates the ACQuire (measurement) operation.
Parameter
Typical Return
(none)
(none)
Initiate a measurement: INIT:ACQ
If the ACQuire operation is already initiated, this command will produce error -213,"Init ignored".
INITiate[:IMMediate]:TRANsient
Initiates the TRANsient operation.
Parameter
Typical Return
(none)
(none)
Initiate a TRANsient operation:
VOLT:MODE STEP
FREQ:MODE STEP
INIT:TRAN
If the TRANsient operation is already initiated, this command will produce error -213,"Init ignored".
INITiate:CONTinuous:ACQuire ON|1|OFF|0
Initiates a new measurement and enables or disables "continuous" mode.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Initiate continuous acquisition mode: INIT:CONT:ACQ ON
If continuous mode is on, acquired data is automatically refreshed without the need for an INIT.
You should normally leave continuous mode off (default).
If the TRANsient operation is already initiated, this command will produce error -213,"Init ignored".
The
and
commands set this parameter to 0.
Keysight AC6800B Series Operating and Service Guide
139
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...