TRIGger:SYNChronize:SOURce IMMediate|PHASe
TRIGger:SYNChronize:SOURce?
Sets the output-on phase control when OUTPut ON is sent.
Parameter
Typical Return
IMMediate|PHASe IMM or PHAS
Enable output-on phase control:
TRIG:SYNC:SOUR PHASE
This parameter is set to IMMediate at power-on or after *RCL or *RST.
TRIGger:SYNChronize:PHASe[:ON] <
value
>|MINimum|MAXimum
TRIGger:SYNChronize:PHASe? [MINimum|MAXimum]
Sets the phase angle of the output-on phase control in degrees.
Parameter
Typical Return
An integer from 0 to 359 d1.02
Set the OUTPUT on phase angle to 180°:
TRIG:SYNC:PHAS 180
This command supports the parameter suffix DEG (degrees), but not RAD (radians).
The
command sets this parameter to 0.
TRIGger:TRANsient[:IMMediate]
Triggers the TRANsient subsystem.
Parameter
Typical Return
(none)
(none)
Configure and trigger the TRANsient subsystem:
VOLT:MODE STEP
FREQ:MODE STEP
TRIG:TRAN:SOUR BUS
INIT:TRAN
TRIG:TRAN
Keysight AC6800B Series Operating and Service Guide
157
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...