[SOURce:]VOLTage:OFFSet:LIMit:LOWer <
value
>|MINimum|MAXimum
[SOURce:]VOLTage:OFFSet:LIMit:LOWer? [MINimum|MAXimum]
[SOURce:]VOLTage:OFFSet:LIMit:UPPer <
value
>|MINimum|MAXimum
[SOURce:]VOLTage:OFFSet:LIMit:UPPer? [MINimum|MAXimum]
Sets the lower and upper DC voltage soft limits.
Parameter
Typical Return
155 V range: -222.5 to 222.5 V
310 V range: -445.0 to 445.0 V
(none)
Set the voltage limits to the be 100 and 350 V:
VOLT:OFFS:LIM:LOW 100
VOLT:OFFS:LIM:UPP 350
This command takes one or three arguments. You cannot have a low or high limit without the other.
This command applies to DC and AC+DC.
Error m166 is generated if the lower limit is beyond the MIN/MAX limits, and error message
+167 is generated if the upper limit is beyond the MIN/MAX limits.
Voltage units (MV, V, KV) are supported.
[SOURce:]VOLTage:OFFSet:MODE FIXed|STEP
[SOURce:]VOLTage:OFFSet:MODE?
Sets the trigger transient mode for DC voltage settings. Specifying FIX (default) disables the trigger
function; specifying STEP enables it.
Parameter
Typical Return
FIXed|STEP
FIX
Set the voltage offset mode to FIXed
VOLT:OFF:MODE FIXed
The
command sets this parameter to FIXed.
Keysight AC6800B Series Operating and Service Guide
163
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...