Keysight B1507A User’s Guide, Edition 1
2- 17
Installation
Connecting Accessories
3-pin Inline Package Socket Module
Use the socket module to connect a three-terminal inline packaged device. See the
following simple instruction.
Requirements
•
3-pin inline package socket module
•
DUT
Procedure
1. If the earthing wire is not connected to the socket module, connect it.
2. Attach the socket module directly to the Base/Gate, Collector/Drain, and
Emitter/Source terminals of the test fixture.
3. Connect the earthing wire of the socket module to the chassis common terminal.
4. Set your DUT on the socket.
5. Close the fixture cover and perform measurement.
WARNING
Hazardous voltage, instrument maximum output voltage may appear at the
measurement terminals if the fixture cover is closed.
WARNING
Open the Guard terminal of the test fixture. Do not connect anything to it.
NOTE
Short plate is furnished with the socket module. It is used for performing the short
correction before the capacitance measurement. Set it on the socket before the short
correction and remove it after the correction.
Socket module
Test fixture
DUT
Chassis
common