Keysight B1542A User’s Guide, Edition 10
6-21
PLSDIV TIS Commands
Command Reference
PlsdivSetPulseWidth
This command sets the pulse width of the gate voltage.
Filename
PlsdivSetPulseWidth.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameter.
pulse_width
Pulse width of gate pulse, in second. Numeric value, 10 ns to 1
s. See
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Figure 6-6
Pulse Setup Parameters
PlsdivSetSampling
This command sets the oscilloscope sampling rate and sampling number for the
pulsed IV measurement.
Remarks
Adjust the sampling number and the sampling rate carefully to obtain the waveform
you desire. To obtain faster measurement throughput, decrease the sampling
number. Too small sampling numbers will cause measurement errors.
Filename
PlsdivSetSampling.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Pulse period
V/2
V/2
PulseWidth
V/10
V/10
TransTime
PulseBase
Vg
V: Pulse amplitude, > 5.5 mV, < 4.5 V
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......