Keysight B1542A User’s Guide, Edition 10
2-7
Installation
Hardware Installation
Connecting Interface Cables
Connect the instruments via GPIB or LAN as shown below.
Requirements:
•
GPIB cable, maximum 4 ea. Quantity depends on your system components
(instruments and computer).
•
LAN cable, 2 ea. only for using the DSO/MSO 9000 series or S series
oscilloscope
Procedure:
1. Connect a GPIB cable between the DC source monitor and the pulse generator.
2. Connect a GPIB cable between the switch controller and one of above
instruments.
3. For the Desktop EasyEXPERT users:
Connect a GPIB cable between your computer and one of above instruments.
4. Connect the oscilloscope as shown below.
•
For using the DSO/MSO 9000 series or S series oscilloscope:
a. Connect the B1500A or your computer to a LAN.
b. Connect the oscilloscope to the LAN.
Subnet mask must be the same for their network settings.
•
For using the other oscilloscope:
Connect a GPIB cable between the oscilloscope and one of above
instruments or computer.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......