3-4
Keysight B1542A User’s Guide, Edition 10
Performing System Setup and Compensation
Pulsed IV System Setup
Table 3-1
Required Accessories to Perform System Setup
Table 3-2
Pulsed IV System Setup
Accessory
Quantity
Triaxial cable
2
BNC cable
1
BNC(m)-Triax(f) adapter, floating guard
1
BNC(m)-(f)-(f) adapter
1
GNDU-chassis adapter
1
When the action must be executed
Action
after system installation
System Configuration
Compensation
after changing interface address of an instrument
System Configuration
after changing a channel assignment
System Configuration
Compensation
after repair of oscilloscope
Compensation
after calibration of oscilloscope
Compensation
after replacing oscilloscope
Compensation
after replacing bias-T
Drain Cable Replacement
after replacing measurement cables to drain
terminal
Drain Cable Replacement
before starting the pulsed IV measurements with
the
VgAdjust
=
Disable
condition
Pgu Compensation
if the skew measurement error occurs
Skew Measurement
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......