Keysight B1542A User’s Guide, Edition 10
5-9
PLSDIV Test Definitions
PLSDIV DC IdVd
Extended Test
Parameters
The following parameters are available to set the test condition.
IgLimit
Gate current compliance, in A. 1 nA to 100 mA, 3 digits or 1 nA
resolution. Set the value that covers current to the terminator.
For example, the current is 100 mA for TermR=50
with
Vg=5 V.
IntegTime
Integration time. SHORT, MEDIUM, or LONG.
HoldTime
Hold time, in s. 0 to 655.35. 5 digits or 10 ms resolution.
DelayTime
Delay time, in s. 0 to 655.35. 5 digits or 100
s resolution.
DcSwitch
Pulse/dc switch units. Yes (exists, DC IV measurement
connection will be automatically made) or No (none).
SWAddress
GPIB address of the switch controller. Integer, 1 to 32.
Test Output
The following output parameters are displayed after the test.
VgSet
Gate voltage, in V. SMU output value.
Vg
Gate voltage, in V. Target value at the gate terminal.
Vdrain
Drain voltage, in V. SMU output value.
Id
Drain current, in A.
IdPerWg
Drain current per unit gate width, in A/
m. IdPerWg = Id / Wg
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......