Keysight B1542A User’s Guide, Edition 10
5-11
PLSDIV Test Definitions
PLSDIV DC IdVd SMU
VdStop
Primary sweep stop voltage for the drain terminal.
-
10 to 10 V,
4 digits or 10 mV resolution.
VdStep
Primary sweep step voltage for the drain terminal.
-
10 to 10 V,
5 digits or 1 mV resolution.
•
VdStart, VdStop, VdStep
The parameters set the primary sweep source used to apply the drain voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VdStep=0, the measurement voltage is
VdStart only.
Number of sweep points = (VdStop
-
VdStart) / 1
•
VgStart, VgStop, VgStep
The parameters set the secondary sweep source used to apply the gate voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VgStep=0, the measurement voltage is
VgStart only.
Number of sweep points = (VgStop
-
VgStart) / 1
•
Gate voltage
To apply the specified voltage (Vg) to the gate terminal, the SMU output voltage
(VgSet) will be set to the following value.
VgSet = Vg * (TermR+CableR) / TermR
Extended Test
Parameters
The following parameters are available to set the test condition.
IgLimit
Gate current compliance, in A. 1 nA to 100 mA, 3 digits or 1 nA
resolution. Set the value that covers current to the terminator.
For example, the current is 100 mA for TermR=50
with
Vg=5 V.
Ranging
Current measurement ranging mode. Auto, LimitedAuto, or
Fixed.
Irange
Current measurement range. 10 pA to 100 mA. 6 digits or 10
pA resolution. Available measurement range depends on the
type of SMU.
ADCType
Only for the E5260/E5270 series. Type of A/D converter.
HighSpeed or HighResolution
IntegMode
Only for the E5260/E5270 series. Integration mode. Auto,
Manual, or PLC.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......