5-14
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
PLSDIV DC IdVg
•
VgStart, VgStop, VgStep
The parameters set the primary sweep source used to apply the gate voltage. The
number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VgStep=0, the measurement voltage is
VgStart only.
Number of sweep points = (VgStop
-
VgStart) / 1
•
VdStart, VdStop, VdStep
The parameters set the secondary sweep source used to apply the drain voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VdStep=0, the measurement voltage is
VdStart only.
Number of sweep points = (VdStop
-
VdStart) / 1
•
Gate voltage
To apply the specified voltage (Vg) to the gate terminal, the SMU output voltage
(VgSet) will be set to the following value.
VgSet = Vg * (TermR+CableR) / TermR
Summary of Contents for B1542A Pulsed IV Package