5-16
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
PLSDIV DC IdVg SMU
PLSDIV DC IdVg SMU
This test definition controls SMUs to apply the drain voltage and the gate voltage,
measures the drain current, and displays the Id-Vg characteristics.
Device Parameters
The following parameters are available to set the test condition.
Polarity
Polarity of source output. Nch (applies the specified value) or
Pch (applies the negative specified value).
Lg
Gate length, in m.
1 nm, 4 digits or 1 nm resolution.
Wg
Gate width, in m.
1 nm, 4 digits or 1 nm resolution.
Temp
Temperature, in degree. 4 digits or 0.001 degree resolution.
IdMax
Maximum drain current measurement range, in A. 0 to 80 mA,
4 digits or 1
A resolution.
Test Parameters
The following parameters are available to set the source output.
CableR
Residual resistance of the gate measurement path from SMU to
terminator.
0
, 4 digits or 1 m
resolution.
TermR
Resistance of terminator.
1 m
, 4 digits or 1 m
resolution.
TermR=No applies 1E+30
.
Gate
Channel number of the SMU connected to the gate terminal.
Slot1 to Slot8. Maximum channel number depends on the
configuration of DC source monitor.
VgStart
Primary sweep start voltage for the gate terminal.
-
10 to 10 V,
4 digits or 10 mV resolution.
VgStop
Primary sweep stop voltage for the gate terminal.
-
10 to 10 V,
4 digits or 10 mV resolution.
VgStep
Primary sweep step voltage for the gate terminal.
-
10 to 10 V,
5 digits or 1 mV resolution.
Drain
Channel number of the SMU connected to the drain terminal.
Slot1 to Slot8. Maximum channel number depends on the
configuration of DC source monitor.
VdStart
Secondary sweep start voltage for the drain terminal.
-
10 to 10
V, 4 digits or 10 mV resolution.
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......