5-36
Keysight B1542A User’s Guide, Edition 10
PLSDIV Test Definitions
Parameters
MeasTime
MeasTime defines the data extraction timing (see
) which is the timing to
extract the pulsed IV measurement data (gate voltage, drain voltage, and drain
current) from the oscilloscope measurement result (pulse waveform). Initially, this
value is set to 0.5. Changing this value may cause a measurement error. If you
change this value, check the pulse waveform and choose the point where the pulse
voltage is flat and stable.
Figure 5-9
MeasTime
SmplNum
SmplNum defines the sampling number set to the oscilloscope. Set the value in
order to display the pulse waveform on the oscilloscope screen. A small number will
improve measurement throughput.
Sampling rate of the oscilloscope is set to the value automatically calculated from
the specified pulse width and the specified sampling number.
offset = 0
offset = 0
Gate pulse
Drain pulse
Data extraction timing
meas_time
x Pulse width
V/2
V/2
Pulse width
V: Pulse amplitude, > 5.5 mV, < 4.5 V
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......