Keysight B1542A User’s Guide, Edition 10
6-19
PLSDIV TIS Commands
Command Reference
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivSetDeviceDelay
This command sets the device delay time between the gate pulse and the drain pulse.
Before performing the pulsed IV measurement, figure out the device delay time and
set it by using this command. To figure out the device delay time, set up the test
system properly, connect the device under test, apply voltage pulse to the gate
terminal, and monitor the gate pulse and the drain pulse as shown in
.
Filename
PlsdivSetDeviceDelay.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameter.
device_delay
Delay time caused by test device, in second. Numeric value,
-
0.000001 to 0.000001.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
Figure 6-4
Adjustment for Device Delay
After adjustment
Before adjustment
Drain monitor pulse
Device Delay
Gate monitor pulse
Drain monitor pulse
Gate monitor pulse
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......