We know you can’t afford instrument downtime due to hardware 
failures and unscheduled maintenance. That’s why our engineers 
designed reliability into the DAQ970A: A rugged enclosure, 
state-of-the-art surface mount construction throughout, reduced 
parts counts, and rigorous and thorough testing on all aspects of 
the product.

Take the guesswork out of relay maintenance

The DAQ970A uses our proprietary relay maintenance system to 
help you to predict relay end-of-life and avoid costly production 
line downtime. It automatically counts every individual switch 
closure and stores it in nonvolatile memory on each module. You 
can query the total number of cycles on any individual channel 
so you can schedule maintenance and avoid erratic end-of-life 
failures.

Spec interpretation guide

The following pages list the technical specifications for the 
Keysight DAQ970A Data Acquisition/System and its modules. The 
explanations and examples below are helpful in understanding 
how to interpret these specifications:

 

– Measurement accuracy is specified as percent of reading plus 

percent of range, where reading is the actual measured value 
and range is the name of the scale (1 V, 10 V, etc.)—not the 
full scale value (1.2 V, 12 V, etc.).

 

–  DMM measurement accuracies include all switching errors. 

Switching errors are also listed separately in the module 
specifications section. Temperature measurement accuracies 
include ITS-90 conversion errors. The thermocouple 
accuracies include the reference junction error as well.

 

– Accuracies are listed as either 24-hour, 90-day, or 1-year 

specifications. This refers to the length of time since the 
instrument’s last calibration. Use the specification that 
matches your calibration cycle. The 24-hour specifications 
are useful for determining short-term relative performance.

Example 1: Basic dcV accuracy

Calculate the accuracy of the following measurement:

9 V dc input
10 V dc range
1-year accuracy specifications
Normal operating temperature (18 - 28°C)

From the following page, the 1-year accuracy is:
0.003% of r 0.0004% or range

 

Which translates into: 

 

(0.003/100 x 9 V)+ (0.0004/100 x 10 V) = 310 μV 

 

For a total accuracy of:

 

310 μV/9 V = 0.0034%

Example 2: Extreme operating temperature

When the DAQ970A is used outside of its 18 - 28°C temperature 
range, there are additional temperature drift errors to consider. 
Assume the same conditions in Example 1, but at a 35°C operat-
ing temperature.

The basic accuracy is again:

0.003% of r 0.0004% of range = 310 μV

Now, multiply the 10 V temperature coefficient from the following 
page by the number of degrees outside of operating range for 
additional error (non ACAL):
(0.0005% r 0.0001% range)
 

/°C x (35 - 28°C) =

(0.0005% r 0.0001% range)
 

/°C x 7°C =

0.0035% r 0.0007% range = 385 µV
 

Total error is then:

 

310 μV + 385 μV = 695 μV or 0.0077%

www.allice.de 

Allice Messtechnik GmbH 

Summary of Contents for DAQ970A

Page 1: ...T E C HNI C A L O V E R V IE W Keysight DAQ970A Data Acquisition System www allice de Allice Messtechnik GmbH...

Page 2: ...tioning and modular flexibility all in a low cost compact data acquisition package The DAQ970A features 6 digits 22 bits of resolution 0 004 basic dcV accuracy and ultra low reading noise Combine that...

Page 3: ...tection and LAN lock out are also provided to limit access The graphical Web interface makes it easy to configure measurements set up and execute scans or trouble shoot your designs from anywhere on t...

Page 4: ...tiation of the measurements 6 Measurement configuration menu to set parameters for measurements 7 Knob 8 Cursor navigation keypad Label Description 1 Slot identifier 100 200 300 2 External trigger inp...

Page 5: ...A is easy to use for a multitude of data logging and monitoring applications either stand alone or with a computer Its flexible modular design makes it scalable from 20 to 120 channels and lets you ad...

Page 6: ...r later uploading to a PC Or if you need even more memory the DAQ970A USB port can be used to log data directly to a USB flash drive or to copy the data from the reading memory without being connected...

Page 7: ...are drivers that support C C Visual Basic Visual Studio LAN and USB on the DAQ970A and SCPI programming language make integration even easier The functionality you need We put a lot of thought into de...

Page 8: ...rement switching and control capabilities to choose from Modules include both low frequency and RF multiplexers a matrix switch a general purpose switch and a multifunction module that includes digita...

Page 9: ...ing errors are also listed separately in the module specifications section Temperature measurement accuracies include ITS 90 conversion errors The thermocouple accuracies include the reference junctio...

Page 10: ...0 0006 0 0040 0 0006 0 0055 0 0006 0 0005 0 0001 0 0002 0 0001 300 V 0 0020 0 0020 0 0035 0 0020 0 0040 0 0020 0 0055 0 0020 0 0005 0 0001 0 0002 0 0001 Resistance2 7 100 0 0030 0 0050 0 0050 0 0060 0...

Page 11: ...Hz 10 Hz 0 10 0 05 0 10 0 06 0 10 0 06 0 11 0 06 0 010 0 008 10 Hz 20 kHz 0 02 0 05 0 04 0 06 0 05 0 06 0 06 0 06 0 010 0 008 20 kHz 50 kHz 0 05 0 09 0 06 0 09 0 07 0 09 0 08 0 09 0 010 0 0012 50 kHz...

Page 12: ...uple DAQ970A extended temp ranges Probe5 Ref junction DMM accuracy E 200 C to 150 C 1 4 C J 210 C to 150 C 1 6 C K 200 C to 100 C 1 7 C N 200 C to 100 C 2 C R 50 C to 300 C 2 9 C T 200 C to 100 C 1 7...

Page 13: ...20 pF 50 pF Capacitance Lo Earth NA 150 pF 135 pF 80 pF 80 pF NA 80 pF Volt Hertz limit 10 8 10 8 10 8 10 8 10 8 10 10 10 8 Characteristics typical Other T C CRJ accuracy Switch life no load Unlimited...

Page 14: ...ms max alarm sampling Latency 5 ms typical to DAQ970A alarm output Read write speed 95 s Max count 2 26 1 Totalize input 100 kHz max Rising or falling edge programmable min H or L time 5uS Signal leve...

Page 15: ...N 275 275 907A Totalizer 240 240 Characteristics Nominal Data out of Memory FETCh of 50K readings Single channel DAQ970A over USB DAQ970A over LAN Readings 55000 s 120000 s Rdgs with timestamp 35000 s...

Page 16: ...t Burden Voltage at Full Scale DC current range Burden voltage 1 A 0 0011 V 10 A 0 011 V 100 A 0 11 V 1 mA 0 11 V 10 mA 0 027 V 100 mA 0 27 V 1 A 0 7 V Characteristics Nominal Other A D linearity 0 00...

Page 17: ...ion EN 61010 2 030 2010 1st Edition ANSI ISA 61010 2 030 82 02 03 First Edition ANSI UL 61010 2 030 First Edition CAN CSA C22 2 No 61010 2 030 First Edition Refer to Declaration of Conformity for curr...

Page 18: ...rature Operating all cards 0 to 55 C all cards 40 to 70 C all cards 40 C to 80 RH Storage Humidity non condensing 1 Not recommended for connection to ac line without external transient suppression Up...

Page 19: ...n 120 V switching The DAQM900A module is divided into two banks of 10 two wire channels each All 20 channels switch both HI and LO inputs thus providing fully isolated inputs to the internal DMM or an...

Page 20: ...ls 22 Com Current Fuse Fuse Internal DMM input 4W sense Internal DMM input Current Reference junction sensor Channel switches Bank switch Shunt switches DAQM902A 16 Channel High Speed Multiplexer Scan...

Page 21: ...rnal measurements except current are supported The module low connection is isolated from earth and can float up to 300 V DAQM903A 20 Channel Actuator General Purpose Switch SPDT Form C latching relay...

Page 22: ...el RF Multiplexer 50 2 GHz bandwidth BNC to SMB adapter cables included The Keysight DAQM905A RF multiplexers offer broadband switching capabilities for high frequency and pulsed signals Use them to r...

Page 23: ...allows great flexibility for a variety of sense and control applications It combines two 8 bit ports of digital input and output a 100 kHz gated totalizer and two 12 V or 24 mA analog outputs all on a...

Page 24: ...e With 26 bits of resolution it can count events at full speed for nearly 11 minutes without an overflow Max count 226 1 Totalize input 100 kHz max Rising or falling edge programmable Signal level 1 V...

Page 25: ...mensions To rack mount two instruments side by side order thelock link kit part number DAQA194A FG and flange kit part number DAQA191A FG To rack mount a single instrument order the adapter kit part n...

Page 26: ...nnel Solid state multiplexer DAQM901A 20 Channel armature multiplexer DAQM902A 16 Channel reed multiplexer DAQM903A 20 Channel actuator general purpose switch DAQM904A 4 x 8 Two wire matrix switch DAQ...

Reviews: