Calibration 3
Series E4360 Service Guide
37
Calibration Procedure
Enter Calibration mode
Front Panel:
SCPI Command:
Select
System\Admin\Login
.
Enter your password in the
Password field and press Select.
CAL:STAT ON, <password>
Voltage Programming, Measurement, and Protection Calibration
Step 1. Connect the DMM’s voltage input to an output channel. Connect the
+sense terminal to the +output and the -sense terminal to the -output.
Step 2. Select the first voltage calibration point.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P1 data
CAL:VOLT:LEV P1
*OPC?
Step 3. Measure the output voltage with the DMM and enter the data.
Front Panel:
SCPI Command:
Select the Measured Data field.
Enter the data from the external
DMM.
Press Select when done.
CAL:VOLT <data>
Step 4. Select the second voltage calibration point.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P2 data
CAL:VOLT:LEV P2
*OPC?
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
SCPI Command:
Select the Measured Data field.
Enter the data from the external
DMM.
Press Select when done.
CAL:VOLT <data>
Step 6. Select voltage protection calibration. Note that the output voltage will go
to near full-scale during this procedure.
Front Panel:
SCPI Command:
Select OVP. Press Select when
done.
Press Next to finish calibration
CAL:VOLT:PROT
*OPC?