Calibration 3
Series E4360 Service Guide
39
Save and Exit Calibration Mode
CAUTION
Storing calibration constants overwrites the existing ones in non-
volatile memory. If you are not sure you want to permanently store the
new constants, do not Save the data when you exit the calibration
mode. The calibration will then remain unchanged.
Front Panel:
SCPI Command:
Select
System\Admin\Cal\Date
.
Enter the calibration date in the
Date field. If desired, you can
enter alphanumeric data in this
field.
Select
System\Admin\Cal\Save
.
Select Save to save all calibration
data.
Select
System\Admin\Logout
to
exit calibration mode.
To enter a calibration date:
CAL:DATE “<date>”, (@1)
To save calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
Change the Admin/Calibration Password
The password must be numeric, and can be up to 15 digits in length.
After it has been changed, you can only enter the calibration mode by
providing the correct password. If the password is lost, access can be
restored by setting an internal switch that resets the password to 0. If
the message “Locked out by internal switch setting” or “Calibration is
inhibited by switch setting” appears, the internal switch is set to prevent
the password from being changed (Refer to “Accessing the Calibration
switch” in chapter 4).
Front Panel:
SCPI Command:
Select
System\Admin\Login
.
Enter the original password in the
Password field and press Select.
Select
System\Admin\Password
.
Enter a new password in the
Password
field.
Logout of the Admin menu to
activate the password.
Enter calibration mode using
the original password
CAL:STAT ON, <password>
To change the password:
CAL:PASS <NRf>
To exit calibration mode and
activate the password:
CAL:STAT OFF