5 Troubleshooting
58
Series E4360 Service Guide
Calibration and Passwords
Calibration is not required after the mainframe has been repaired. All
calibration information is module-specific and resides in each output
module.
However, if the A6 board that contains the calibration switches has been
replaced, any calibration password that has been set will be lost.
Passwords that are used to prevent access to the front panel Admin
menu will also be lost.
The calibration password will automatically be reset to the factory
default, which is zero (0). Front panel access to the Admin menu will be
unrestricted unless a password is set.
Refer to the User’s Guide for information on setting the calibration
password as well as the Admin menu password.
Initialization
The solar array simulator's model number, serial number, and other
constants that are required to program the instrument are stored in an
EEPROM on the A6 board. If this board is replaced, the instrument must
be re-initialized.
To initialize the solar array simulator, you must send the following
information to the solar array simulator over the interface:
DIAGnostic:FRAMe:ATTRibute "ModelNumber", "value"
where “value” is the model number, e.g. E4360B.
DIAGnostic:FRAMe:ATTRibute "SerialNumber", "value"
where “value” is the serial number located on top of the interface cover.
At power-up, the solar array simulator will go through the turn-on self
test sequence. All other parameters will be set to the factory defaults.
The mainframe is now initialized.