Appendix A Error Messages
68
Series E4360 Service Guide
Device-dependent Errors
(continued)
313
Datalog time interval out of range
The datalog time interval is set to a value that is outside the allowable range.
315
Settings conflict error
There is a settings conflict error. For example, the Vmp setting may be greater than the Voc setting.
This is not allowed.
316
Mass storage error
An error has occurred with the mass storage. For example, the mass storage could be full.
318
Datalog error
An error has occurred with the data logger.
320
VMP must be less than VOC
The value of the maximum-point voltage must be less than the value of the open-circuit voltage.
321
IMP must be less than or equal to ISC
The maximum-point current value must be less than or equal to the short-circuit current value.
322
VMP and/or IMP too small
The values of either the maximum-point voltage or the maximum-point current are too small.
323
Can’t init unless output in SAS mode with List submode selected
The instrument cannot be initialized unless it is set to SAS mode with List mode selected.
324
Cannot change resolution unless in Fixed mode
The instrument must be set to Fixed mode operation before the resolution can be changed.
325
Too many table points
Too many table points have been specified. The maximum is 4000.
326
Table not found
A table with that name has not been found.
327
Cannot de-select table in Table mode
The table cannot be de-selected because the instrument is operating in Table mode. Set the
instrument to Fixed mode to de-select the table.
328
Calculated open circuit voltage too large
The calculated open-circuit voltage exceeds the output voltage rating of the instrument.
329
Exceeds soft limit
The programmed value exceeds the soft limit setting.
330
Table error
An error has occurred because of one of the table value parameters
.
331
Exceeds low-line limit
The programmed current exceeds the reduced current that is available when the instrument is
operating at low AC line (100 VAC or 120 VAC nominal).
400
Power must be cycled for new settings to take effect
Cycle AC power for the new settings to take effect.
401
IO settings have been reset to factory defaults. Cycle power for settings to take effect
Cycle AC power for the new IO settings to take effect.
402
NVRAM settings have been reset to factory defaults. Cycle power for settings to take effect
Cycle AC power for the new NVRAM settings to take effect.
414
Settings conflict warning
Settings conflict error. For example, the Vmp setting may be greater than the Voc setting.
Command Errors
(these errors set Standard Event Status register bit #5)
−
100
Command error
Generic syntax error.
−
101
Invalid character
An invalid character was found in the command string.
−
102
Syntax error
Invalid syntax was found in the command string. Check for blank spaces.