30 Keysight E5053A Microwave Downconverter
Connecting Devices to E5052B for Operation
Basic Operation
3-
Turning on Power
After connecting the E5052B and the E5053A, follow these steps to turn on the
power.
Step 1.
Turn on the E5053A.
Step 2.
Turn on the E5052B.
Basic Operation
For information on operation, see the "Measurement Using the E5053A and a
External Mixer" chapter in the "E5052B Signal Source Analyzer User's Guide."
If the E5052B is turned on first, the E5053A cannot be recognized, and the
downconverter setting of the E5052B is disabled. Note that, if the auto
recall function with the function of the E5053A enabled is executed under
this condition, recall is performed normally, but the “No Downconverter
Unit Connected” error message appears. They become available by
enabling the downconverter setting manually.