Using Wideband IF Output
The following steps demonstrates how to use the wideband block converted IF
signal from the N9041B’s millimeter wave downconverter for analysis on the
DSOS804 oscilloscope and VSA89600 software.
Step
Action
Notes
1
Connect the N9041B and DSOS804
oscilloscope.
a
Connect the
Ext IF Out
port from the
N9041B to the oscilloscope’s
Ch1
input as shown in
below.
b
Connect the
10 MHz OUT
port from
the N9041B to the
10 MHz Ref In
of
the DSOS804.
Figure 11
Connection diagram between the N9041B and DSOS804
Note that the N9041B-CRW option is required for the Ext IF Out connector.
2
Set spectrum analyzer in IQ Analyzer
mode.
a
Press the
MODE/MEAS
hard key or tap
the
Screen
tab.
b
For Mode, Measurement, and View
choose
IQ Analyzer
,
Complex
Spectrum
, and
Normal
.
The scope requires time sampled
waveform, so the mode must be changed
to IQ Analyzer mode.