4. STEP 3: Turn On Chassis and
Verify Operation
Do not install modules in the peripheral slots yet—this is done in
STEP 4. The only module that should be installed in the chassis
at this time is either an embedded controller or a PXIe system
module in slot 1 as described in
This guide assumes that chassis preparation, turn-on and
verification are performed on the bench prior to installing the
chassis in a rack. If you prefer to install the chassis in a rack first,
see the
on the Software
Keysight PXIe Chassis Family User Guide
and Product Information CD for rack mounting information.
If you encounter difficulty in turning on the chassis system, see
the troubleshooting information in the
Keysight PXIe Chassis
.
Family User Guide
This instrument has auto-ranging power supplies, meaning that the chassis
does not have a voltage selector switch. Ensure that the AC supply line voltage
is within the range of 100-120 VAC or 220-240 VAC, and 50/60 Hz frequency.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal condition
(in which all means for protection are intact) only.
When you press the chassis power-on button, if the chassis does not power up
and the front panel LEDs do not light, it is possible for the chassis to be in a
safety shutdown state. Remove the chassis AC power cord from the chassis for
one minute. Reconnect the power cord and turn on the chassis again. If it still
does not power on, refer to the troubleshooting information in the
Keysight
.
PXIe Chassis Family User Guide
Power sequencing of the chassis
When powering up the system, the chassis should be powered up first. After
powering up the chassis, you should wait at least three seconds before turning on
the PC. The chassis front panel temperature LED, which is on for three seconds
after the chassis is powered up, provides a convenient way to measure this delay,
as shown below.