M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
25
Chassis Installation
2
Installing and Removing AXIe Modules
While the AXIe chassis accepts all modules conforming to the single slot, 1U AXIe
standard, option BC1 support only M9703B, M979A, and M97010A digitizer
modules. Other modules, including AXIe embedded controller modules, may be
installed, but they will not function while the integrated controller is enabled. AXIe
Filler Modules may be installed, but are NOT shipped with the chassis. AXIe Filler
Modules are order-able separately with Keysight part number Y1221A.
The insertion and removal procedures are the same for all module types except
filler modules.
CAUTION
Static Electricity
—The components and connectors on modules are
sensitive to static electricity. To minimize electrostatic damage,
take the necessary anti-static precautions.
Empty Slots
—Except for performing initial chassis verification or
troubleshooting, do not operate the chassis with empty slots.
Always insert a filler module or instrument module into empty
slots. This is especially important for the slots on either side of an
instrument module. This allows proper air flow and cooling, and
provides EMI shielding for the chassis and installed components.
Leaving slots empty can increase fan speed, raise ambient noise,
overheat components, and shut down modules.
ESM
—The Embedded System Manager is integral to the operation
of the chassis. Except for troubleshooting purposes, do not
remove the ESM.
Hot Swap
—AXIe does not explicitly support hot swap for instrument
modules. Keysight recommends fully powering down the chassis
before installing or removing modules.