2
Chassis Installation
28
M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
7
Tighten the captive retaining screws at both sides of the module.
.
8
Retract the handle ends by sliding them outward on their metal shafts, away
from each other, toward the chassis edge; this secures them out of the way of
test connections.
9
Repeat steps
through
for additional modules, as needed. Ensure that each
slot has an instrument or AXIe Filler Module (Y1221A) installed.
10
Power up the AXIe chassis. Verify that the chassis fans are operating and free of
obstructions that may restrict airflow.
Removing a Module
The instructions below apply to all module types. The Embedded System Module
(ESM) has the same extraction handles and retaining screws as instrument
modules. If you should ever have to remove the ESM, follow the instructions for
instrument modules below.
1
Power down but do not unplug the AXIe chassis from the AC power.
2
Fully loosen the captive retaining screws on both sides of the module.
3
For an AXIe Filler Module, grasp the panel by the two captive retaining screws,
and slide it out of the chassis.
For all other modules, locate the insertion/extraction handles at each side of
the module’s front panel. Extend the plastic ends of both handles by sliding
them outward on their metal handle shafts, inwards towards each other.
4
Open the handles by pivoting them out towards you, away from the chassis.
Place each thumb at the inside of the handle, forefinger outside the handle, and
rotate the handles with your thumbs. When the handles are perpendicular with
the chassis, stop. The module should now be unseated the module from the
chassis backplane and its faceplate from the chassis front panel.
5
Grasp the levers to slide the module out of the chassis.
CAUTION
Modules are grounded through the chassis. Tighten the module retaining
screws to ensure a proper ground connection.
CAUTION
Ensure you fully loosened the captive module retaining screws before trying to
extract any module. If you attempt to pull the module out by the screws (for
Filler Modules) or by using the extraction handles (other modules) with these
screws still engaged, damage to the chassis or module could result.