Chassis Installation
2
Mounting the Chassis
Environmental and Power Considerations
Consider your anticipated chassis weight (with modules installed), power and
ventilation requirements, benchtop use and rackmounting options to ensure safety
and optimize access to test and control instrumentation.
Weight
The M9506A chassis weighs 23.5 kg (51.8 lbs) with ESM but no other
installed modules. It weighs significantly more fully loaded with AXIe modules. It is
typically used rackmounted, but can also be used on the bench, with proper
support and ventilation.
AC Input
The chassis power limits are:
The chassis Shelf Manager (ESM) knows whether the power supply is connected
to a Low Line or a High Line AC source and the power requested the installed AXIe
modules. At power-up, if any module requests more power than is available, then
the Shelf Manager prevents one or more modules from powering up.
Use the chassis Web Interface (see
M9506A User Guide
, Chapter 3, “Navigating
the Web Interface”) to determine if all modules are powered on.
100/120 VAC (Low Line), 50-60 Hz
200/240 VAC (High Line), 50-60 Hz
1380 W maximum
2000 W maximum
NOTE
The instrument can operate with mains supply voltage fluctuations
up to
±
10% of the nominal voltage.
NOTE
The main power cord can be used as the system disconnecting
device. It disconnects the mains circuitry from the mains supply.
WARNING
Avoid overloading an electrical circuit. Ensure your AC power
line matches the AC Input requirements listed above.
In case you need to power down the chassis in an emergency,
make sure that you have clear and quick access to the primary
disconnect.
If the chassis is rack-mounted, this primary
disconnect can be a power system on the rack or the circuit
breaker on the chassis.