2 Getting Started
Instrument Physical Characteristics
Height: 89 mm (3.5 in)
Depth: 559 mm (22 in)
Width : 427 mm (16.8 in)
Weight: 17.2 kg (37.9 lb)
Meeting Environmental and Electrical Requirements
To avoid the loss of data or current user instrument states that have not been permanently saved to non-
volatile memory, the instrument should always be powered down either via the instrument's front panel
power button or the appropriate SCPI command. Instrument's installed in rack systems and powered
down with the system rack power switch rather than the instrument's front panel switch display an Error -
310 due to the instrument not being powered down correctly.
Environment
indoor use unless specified otherwise
altitudes < 9,842.5 feet (3,000 meters)
0 to 50°C temperature
maximum relative humidity (non-condensing): Type Tested, 95% relative humidity up to 40°C decreasing linearly
to 45% relative humidity at 50°C.
From 40°C to 50°C, the maximum % Relative Humidity follows the line of constant dew point.
This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC
61010-1 Third Edition and 664 respectively.
Environmental Testing
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to
be robust against the environmental stresses of Storage, Transportation and End-use; those stresses include but are not lim-
ited to temperature, humidity, shock, vibration, altitude and power line conditions. Test Methods are aligned with IEC
60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Ventilation
VENTILATION REQUIREMENTS: When installing the instrument(s) into a cabinet consideration shall be
given to the convection flow into and out of the cabinet. Consideration shall also be given to the indi-
vidual instruments to avoid having the heated discharge of one instrument, now becoming the cooling
intake air for another instrument.
Another area of concern is verification that the maximum ambient operating temperature of the instru-
ment(s) is not exceeded by cabinet installation.
Keysight recommends forced air convection whenever an instrument(s) are installed in a cabinet and fur-
ther recommends that the maximum operating temperature of the cabinet be reduced 10°C from the low-
est, of the maximum operating temperature of a single instrument.
If there are any concerns or special requirements a Keysight Field Engineer should be consulted to assure
instrument(s) temperature compliance and performance.