20/44
Scienlab Regenerative AC Emulator
–
Keysight SL1200A Series
–
Operating Instructions
1 System Description
Real Power Measurement
10
W
2.3% (working point Urms=230V, Irms=30A)
Apparent Power Measurement
10
VA
Reactive Power Measurement
10
VA
Power Factor Measurement
0.01
4.3% (working point Urms=230V,
Irms=30A cosPhi=0.95)
Output Energy Measurement
0.001
kWh
0.084% (working point Urms=230V, Irms=30A)
(SL1201A – SL1203A)
0.079% (working point Urms=230V, Irms=30A)
(SL1212A – SL1213A)
Options
DC Output
DC Output Ratings
SL1201A
SL1202A
SL1203A
SL1212A
SL1213A
Voltage
500
500
500
1000
V
1000
V
Current
4
180
A
180
A
360
A
90
A
180
A
Power
4
40
kW
60
kW
120
kW
60
kW
120
kW
Output phase
DC+, DC-
Measurement value acquisition
Measurement value acquisition (DC)
Resolution
Accuracy
Current measurement
0.1
A
±18
mA ±0.03% (SL1201A, SL1202A, SL1213A)
4
±36
mA ±0.03% (SL1203A)
4
±9
mA ±0.03% (SL1212A)
4
Voltage measurement
0.1
V
±75
mV ±0.03% (SL1201A – SL1203A)
±150
mV ±0.03% (SL1212A – SL1213A)
Supplemental Characteristics
Measurement value acquisition (DC)
Current rise time (10% to 90%)
<2 ms
Voltage dynamic (10% to 90%)
500
V/ms
Load regulation
0.25% FS
4
Specification valid in single-phase mode