33/44
Scienlab Regenerative AC Emulator
–
Keysight SL1200A Series
–
Operating Instructions
Commissioning 4
4
Commissioning
Danger to life in case of misuse at the outputs.
The Scienlab Regenerative AC Emulator does not have its own insulation guard.
If the system emulates an electrical isolation against PE at the output, the lo-
cal installation regulations and guidelines must be observed.
This would be in the case of the emulation of an IT network in AC mode, or
generally in DC mode.
Electrical malfunction through moisture condensation!
In case of high fluctuations of temperature, condensate water can be formed.
Thus, possible insulation sections cannot be maintained, and the electronics are
damaged.
Before switching on the system, the Scienlab Regenerative AC Emulator accli-
matize to the ambient temperature in the operating area (generally between
5 °C and 40 °C) for at least 24 hours.
Hearing protection due to noise exposure in the operating area!
Noise level < 85 dB(A) at a distance of 1 m from the front
It is recommended to wear hearing protection during operation.
The system may only be switched on and put into operation by specialist staff
and specially trained personnel. (see
Qualification of the Operating Personnel
The following cabling must be completed before commissioning:
Electrical installations are complete (see
),
Load and measuring lines are connected with the DUT (see
),
Signal and control lines are connected (see
),
All connections and cabling between SL1200A, control computer /
Hardware-in-the-Loop (HiL) platform and higher-level safety technol-
ogy, if available, have been checked and are according to the enclosed
circuit diagram,
the device under test is in a protected environment, in accordance to
the safety guidelines, especially regarding protection earthing;
all necessary enable signals are provided.
NOTICE
NOTICE
CAUTION
DANGER