Troubleshooting and Fault Elimination 6
6
Troubleshooting and Fault Elimination
Troubleshooting and fault elimination may only be carried out by specially trained
personnel. (see
Qualification of the Operating Personnel
The following overview provides information about faults, their causes and
remedies.
If an overload or a fault current interruption occurs or if a fault condition
exists:
Notify trained service personnel.
If necessary, notify KeysightCare
Contact KeysightCare
Keysight Technologies Technical Support
www.keysight.com/find/keysightcare-contact
Fault message via the PC software
If the signal lamp or PC software
SL1200A Regenerative AC Emulator Soft Front Panel (SFP)
report an error, the error description can be read out via the PC software.
The detailed user manual is included in the scope of delivery.
System cannot be switched on
The system does not switch on:
Possible causes
Remedy
No supply voltage present
Switch on the building supply voltage
Emergency stop push button pressed
Release emergency stop push button
Signal via higher-level emergency stop/release
Check laboratory safety technology
Not properly wired external emergency/release
in/out
Check wiring inside the SL1200A
Connectivity problem
Check cabling
Fuse has tripped
Switch fuse on after opening the panel; reattach
panel afterwards
Table 5:: Possible cause and remedy
Tripping of the Surge Protection
If the surge protection trips, the fuses must be replaced. Only Keysight-
Care staff is allowed to do this.
NOTICE