Safety and Regulatory Notices
Residual voltage after switch-off.
In case of connected DUT, the output terminals can still carry voltages even after
switch-off of the system!
Measure the voltage at the outputs to ensure that they are voltage-free.
Danger to life due to charging or discharging batteries.
Batteries are sensitive to overloading/underloading and might explode or start
burning. This can lead to danger for life and limb.
Therefore, a battery must be monitored by a battery management system
(BMS).
Hearing protection due to noise exposure in the operating area!
Noise level < 85 dB(A) at a distance of 1 m from the front
It is recommended to wear hearing protection during operation.
Safety and Monitoring Devices
The system has monitoring devices that switch-off the system in case of
emergency. If additional systems are connected to the SL1200A Scienlab
Regenerative AC Emulator, it is recommended that the systems are inte-
grated into a higher-level safety device.
Main switch
•
On the front side
•
I ON: power supply switched on
•
O OFF: power supply switched off, communication with the
system is maintained
•
Trip: when
emergency stop is triggered → system switches
off, main switch in position Trip
•
Note: In the 0 OFF position, the main switch can be locked
using a padlock
Emergency stop push button
•
On the front side
•
Press: forces local emergency stop of the system
(Responds: see safety relay)
•
Turn counterclockwise: unlocks
•
Functional test of the emergency stop button: monthly
DANGER
NOTICE
DANGER