26
Keysight E1406A Command Module Service Guide
Verification Tests
Command Module Self-Tests
Example Program
This program performs an GPIB power-on test for the Keysight E1406A
command module and uses the SYST:ERR? command to check results. If the
power-on test passes, +0,"No error" is returned. If the power-on test fails, the
test returns an error message for each error detected. In this case, see Chapter 5
- Error Messages for an explanation of the error(s)
1!Test S-1: GPIB Power-On Self-Test
2!
10CLEAR SCREEN
20 ASSIGN @Addr to 70900
Figure 2-4
Test S-1: GPIB Power-On Test
If the Ready light does not turn on and/or the Failed/SYSFAIL
lights stay lit when power is turned ON, there is a very high
probability that the command module is defective. In this case,
this test will probably not run, and you should see Chapter 4 -
Service for repair/replacement guidelines.
Summary of Contents for VXI bus 75000 C Series
Page 1: ...Keysight E1406A Command Module Service Guide 75000 Series C...
Page 2: ......
Page 6: ...6...
Page 10: ...x Keysight E1406A Command Module Service Guide...
Page 70: ...70 Keysight Keysight E1406A Command Module Service Guide Replaceable Parts Component Locators...
Page 77: ...Keysight E1406A Command Module Service Guide 77 Assembly Disassembly Instructions Service...
Page 78: ...78 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 79: ...Keysight E1406A Command Module Service Guide 79 Assembly Disassembly Instructions Service...
Page 81: ...Keysight E1406A Command Module Service Guide 81 Assembly Disassembly Instructions Service...
Page 82: ...82 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 83: ...Keysight E1406A Command Module Service Guide 83 Assembly Disassembly Instructions Service...
Page 84: ...84 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 106: ...106 KeysightE1406A Command Module Service Guide Error Messages Introduction...
Page 129: ......