KeysightE1406A Command Module Service Guide
91
Introduction
Error Messages
E rror codes read from the error queue are preceded by the num
ber 21. For exam ple, error code 11 displayed on an R S-232 term
inal or printer appears as 2111 when returned by the SYST :ERR?
com m and.
If a fatal error has occurred and the system instrument has not
started, SYST :ERR? cannot be used to read the error queue.
Table 5-2
Language Related Errors
Code
Message
Cause
Command Errors
-
101
INVALID CHARACTER
Unrecognized character in specified param eter.
-
102
SYNTAX ERROR
Com m and is m issing a space or com m a between param eters
-
103
INVALID SEPARATOR
Com m and param eter separated by som e character other than a com m a.
-
104
DATA TYPE ERROR
The wrong data type (i.e. num ber, character, string expression) was used
when specifying a param eter.
-
108
PARAMETER NOT ALLOWED
Param eter specified in a com m and which does not require one.
-
109
MISSING PARAMETER
No param eter specified in com m and in which a param eter is required.
-
113
UNDEFINED HEADER
Com m and header was incorrectly specified.
-
123
NUMERIC OVERFLOW
A param eter specifies a value greater than the com m and allows.
-
128
NUMERIC DATA NOT ALLOWED
A num ber was specified for a param eter when a letter is required.
-
131
INVALID SUFFIX
Param eter suffix incorrectly specified (e.g. .5SECOND rather than .5S).
-
138
SUFFIX NOT ALLOWED
Param eter suffix is specified when one is not allowed.
-
141
INVALID CHARACTER DATA
The discrete param eter specified is not allowed.
-
16
0
BLOCK DATA ERROR
The block sent either contained m ore data than the flash ROM’s could hold or
the block count field disagreed with the num ber of bytes sent.
-
178
EXPRESSION DATA NOT ALLOWED
A param eter other than the channel list is enclosed in parentheses.
Execution Errors
-
211
TRIGGER IGNORED
Trigger occurred from a source other than the specified source.
-
222
DATA OUT OF RANGE
The param eter value specified is too large or too sm all.
-
224
ILLEGAL PARAMETER VALUE
The num eric value specified is not allowed.
-
240
HARDWARE ERROR
Hardware error detected during power-on cycle.
-
252
MISSING MEDIA
No program m able ROM was found, or hardware m alfunction.
Summary of Contents for VXI bus 75000 C Series
Page 1: ...Keysight E1406A Command Module Service Guide 75000 Series C...
Page 2: ......
Page 6: ...6...
Page 10: ...x Keysight E1406A Command Module Service Guide...
Page 70: ...70 Keysight Keysight E1406A Command Module Service Guide Replaceable Parts Component Locators...
Page 77: ...Keysight E1406A Command Module Service Guide 77 Assembly Disassembly Instructions Service...
Page 78: ...78 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 79: ...Keysight E1406A Command Module Service Guide 79 Assembly Disassembly Instructions Service...
Page 81: ...Keysight E1406A Command Module Service Guide 81 Assembly Disassembly Instructions Service...
Page 82: ...82 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 83: ...Keysight E1406A Command Module Service Guide 83 Assembly Disassembly Instructions Service...
Page 84: ...84 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 106: ...106 KeysightE1406A Command Module Service Guide Error Messages Introduction...
Page 129: ......