40
Keysight E1406A Command Module Service Guide
Verification Tests
Test F-3: Hierarchy/Device Information
Set up Equipment·
–
Turn mainframe power OFF
–
Connect computer to command module (see Figure 2-10)
–
Turn mainframe power ON
Device Class
Integer between 0 and 5
0 = VXIbus memory device
1 = VXIbus extended device
2 = VXIbus message based device
3 = VXIbus register based device
4 = hybrid device
5 = Non-VXIbus devic
e
Address Space
Integer from 0 to 15 (Integer value is sum of
binary weighted codes of the address space
occupied by device)
1 = device has A16 registers
2 = device has A24 registers
4 = device has A32 registers
8 = device has A64 registers
A16 Memory Offset
Integer between -1 and 65535
Base address for A16 registers on device.
-1 = device has no A16 memory
A24 Memory Offset
Integer between -1 and 16777215
Base address for A24 registers on device.
-1 = device has no A24 memory
A32 Memory Offset
Integer between -1 and 4294967295
Base address for A32 registers on device.
-1 = device has no A32 memory
A16 Memory Size
Integer between -1 and 65535
Number of bytes reserved for A16 registers.
-1 = device has no A24 memory
A24 Memory Size
Integer between -1 and 16777215
Number of bytes reserved for A24 registers.
-1 = device has no A16 memory
A32 Memory Size
Integer between -1 and 4294967295
Number of bytes reserved for A32 registers.
-1 = device has no A32 memory
Slot Number
Integer between -1 and the number of slots in
mainframe
-1 = slot that contains this device is unknown
Slot 0 Logical Address
Integer between -1 and 255
-1 = Slot 0 device associated with this device is
unknown
Subclass Register
Integer representing the subclass register
contents
-1 = Subclass register not defined for this device
Attribute Register
Integer representing the attribute register
contents
-1 = Attribute register not defined for this device
Manufacturer’s specific
comments
80-character string containing instrument
name and secondary address UNLESS
start-up error(s) are detected.
For start-up errors, return has form “CNFG
ERROR: n, m, ...,z”. See Chapter 5 for error
descriptions.
Summary of Contents for VXI bus 75000 C Series
Page 1: ...Keysight E1406A Command Module Service Guide 75000 Series C...
Page 2: ......
Page 6: ...6...
Page 10: ...x Keysight E1406A Command Module Service Guide...
Page 70: ...70 Keysight Keysight E1406A Command Module Service Guide Replaceable Parts Component Locators...
Page 77: ...Keysight E1406A Command Module Service Guide 77 Assembly Disassembly Instructions Service...
Page 78: ...78 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 79: ...Keysight E1406A Command Module Service Guide 79 Assembly Disassembly Instructions Service...
Page 81: ...Keysight E1406A Command Module Service Guide 81 Assembly Disassembly Instructions Service...
Page 82: ...82 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 83: ...Keysight E1406A Command Module Service Guide 83 Assembly Disassembly Instructions Service...
Page 84: ...84 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Page 106: ...106 KeysightE1406A Command Module Service Guide Error Messages Introduction...
Page 129: ......