Service
1
Keysight 1000B Series Oscilloscopes Service Guide
25
9
Press
[Acquire]
.
10
Press the
Acquisition
softkey until “Average” appears.
11
Press
Average
and turn the entry knob until “8” appears.
12
Press
[Measure]
.
13
Press the
Voltage
softkey.
14
Turn the entry knob to select the
Vpp
menu item.
15
Set the signal generator to a 1 MHz sine wave with a peak-to-peak amplitude
of about 6 divisions as it appears on the oscilloscope screen.
Figure 7
Averages Menu Item