36
Keysight 1000B Series Oscilloscopes Service Guide
1
Service
Channel 2
2 mV/div
±6 mV
+0.96 to +1.04
5 mV/div
±15 mV
+0.96 to +1.04
10 mV/div
±30 mV
+0.97 to +1.03
20 mV/div
±60 mV
+0.97 to +1.03
50 mV/div
±150 mV
+0.97 to +1.03
100 mV/div
±300 mV
+0.97 to +1.03
200 mV/div
±600 mV
+0.97 to +1.03
500 mV/div
±1.5 V
+0.97 to +1.03
1 V/div
±3.0 V
+0.97 to +1.03
2 V/div
±6.0 V
+0.97 to +1.03
5 V/div
±15.0 V
+0.97 to +1.03
10 V/div
±30.0 V
+0.97 to +1.03
Table 13
DC Gain Test (continued)
Vertical
Sensitivity
Power
Supply
Setting
VDMM+
VDMM-
VScope-
Calculated
DC Gain
Offset Gain Test
Limits
Table 14
Analog Bandwidth - Maximum Frequency Check
Vin @ 1 MHz
Vout @
1 MHz
Calculated Gain
@ 1 MHz (Test
Limit = greater
than -3 dB)
Vin @ Max
Freq
Vout @ Max
Freq
Calculated Gain
@ Max Freq (Test
Limit = greater
than -3 dB)
Channel 1
Channel 2
Max frequency: DSO1152B = 150 MHz, DSO1102B = 100 MHz, DSO1072B = 70 MHz, DSO1052B = 50 MHz