106
Keysight 1000B Series Oscilloscopes User’s Guide
4
Making Measurements
Counter (Frequency)
The 1000B Series oscilloscopes have an integrated 6-digit hardware frequency
counter.
The counter operates on the currently selected trigger source and can measure
frequencies from 5 Hz to the bandwidth of the oscilloscope.
The counter uses the trigger comparator to count the number of cycles within a
period of time (known as the gate time), so the trigger level must be set correctly.
The frequency counter is not available in the Alternate trigger mode.
To turn the hardware frequency counter on or off:
1
Press
[Meas]
.
2
In the Measure menu, press
Counter
to toggle the frequency counter display
“ON” or “OFF”.