283
Chapter 5 Error Messages
Execution Errors
4
5
-221
Settings conflict;
amplitude changed due to function
In some cases, the amplitude limits are determined by the output units
currently selected. This may occur when the units are Vrms or dBm due
to the differences in crest factor for the various output functions.
For example, if you output a 5 Vrms square wave (into 50 ohms) and
then change to the sine wave function, the function generator will
automatically adjust the output amplitude to 3.536 Vrms (upper limit for
sine waves in Vrms).
-221
Settings conflict;
offset changed on exit from dc function
In the dc voltage function, the voltage level is controlled by adjusting the
offset voltage (the current amplitude is ignored). When you select a
different function, the function generator will adjust the offset voltage as
needed to be compatible with the current amplitude setting.
-
221
Settings conflict;
FM deviation cannot exceed carrier
The carrier frequency must always be greater than or equal to the
frequency deviation. If you set the carrier to a value less than the
deviation frequency (with FM enabled), the function generator will
automatically adjust the deviation to the maximum value allowed with the
present carrier frequency.
-221
Settings conflict;
FM deviation exceeds max frequency
The sum of the carrier frequency and deviation must be less than or
equal to the maximum frequency for the selected function
plus 100 kHz
(10.1 MHz for sine and square waveforms). If you set the carrier to a
value that is not valid, the function generator will automatically adjust
the deviation to the maximum value allowed with the present carrier
frequency.
-221
Settings conflict;
PWM deviation decreased due to pulse parameters
The PWM deviation is limited by the pulse width or duty cycle, edge
time, and period. The PWM deviation (width or duty cycle deviation) is
adjusted to fit within these limits. See “Pulse Width Modulation (PWM
Commands” in chapter 4 for further information.
33210A users guide.book Page 283 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...