35
Chapter 2 Front-Panel Menu Operation
To Select the Output Termination
4
2
To Select the Output Termination
The Keysight 33210A has a fixed series output impedance of 50 ohms
to the front-panel Output connector. If the actual load impedance is
different than the value specified, the displayed amplitude and offset
levels will be incorrect. The load impedance setting is simply provided
as a convenience to ensure that the displayed voltage matches the
expected load.
1 Press
.
2 Navigate the menu to set the output termination.
Press the
Output Setup
softkey and then select the
Load
softkey.
3 Select the desired output termination.
Use the knob or numeric keypad to select the desired load impedance or
press the
Load
softkey again to choose “High Z”.
To Reset the Function Generator
To reset the instrument to its factory default state, press
and then
select the
Set to Defaults
softkey. Press
YES
to confirm the operation.
For a complete listing of the instrument’s power-on and reset conditions,
see
“Keysight 33210A Factory Default Settings” on page 144
.
33210A users guide.book Page 35 Wednesday, July 16, 2008 11:16 AM
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator User s Guide ...
Page 2: ......
Page 3: ......
Page 14: ...12 Contents Contents 33210A users guide book Page 12 Wednesday July 16 2008 11 16 AM ...
Page 15: ...1 1 Quick Start 33210A users guide book Page 13 Wednesday July 16 2008 11 16 AM ...
Page 53: ...3 3 Features and Functions 33210A users guide book Page 51 Wednesday July 16 2008 11 16 AM ...
Page 273: ...5 5 Error Messages 33210A users guide book Page 271 Wednesday July 16 2008 11 16 AM ...
Page 301: ...6 6 Application Programs 33210A users guide book Page 299 Wednesday July 16 2008 11 16 AM ...
Page 311: ...7 7 Tutorial 33210A users guide book Page 309 Wednesday July 16 2008 11 16 AM ...
Page 335: ...8 8 Specifications 33210A users guide book Page 333 Wednesday July 16 2008 11 16 AM ...