CALCulate:TRANsform:HISTogram:RANGe:AUTO {ON|1|OFF|0}
CALCulate:TRANsform:HISTogram:RANGe:AUTO?
Enables or disables automatic selection of the histogram's lower and upper range values.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON. 0 (OFF) or 1 (ON)
See
l
ON:
the instrument uses the first 1,000 measurements to set the lower and upper range values.
l
OFF:
the lower and upper range values are specified by
CALCulate:TRANsform:HISTogram:RANGe:
l
Setting the lower or upper range value (
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer
) dis-
ables automatic selection of the lower and upper range values (
late:TRANsform:HISTogram:RANGe:AUTO OFF
).
l
The instrument restarts automatic range value selection (if enabled) when
or
is executed.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} {<
value
>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer}? [{MIN|MAX|DEF}]
Sets the histogram's lower and upper range values. Setting the lower or upper range value (
late:TRANsform:HISTogram:RANGe:{LOWer|UPPer
) disables automatic selection of the lower and upper
range values (
CALCulate:TRANsform:HISTogram:RANGe:AUTO OFF
).
Parameter
Typical Return
-1.0E+15 to -1.0E-15,
or
0.0 (default)
or
+1.0E-15 to 1.0E+15
+1.0006
See
.
l
If automatic range value selection is enabled (
CALCulate:TRANsform:HISTogram:RANGe:AUTO ON
),
the query returns the computed range value. If no histogram data exists, 9.91E37 (Not a Number) is
returned.
l
Lower and upper range values are computed automatically if
late:TRANsform:HISTogram:RANGe:AUTO
is ON.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
220
Keysight True
volt
Series Operating and Service Guide