CALCulate:SMOothing[:STATe] {OFF|ON}
CALCulate:SMOothing[:STATe]?
Enables or disables a smoothing (moving average) filter on the measurements. Smoothing is intended to
average out small variations in the measurements. Larger variations will cause the filter to reset.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON.
0 (OFF) or 1 (ON)
Enable a 50-point moving average filter on 1000 measurements:
CONF:VOLT 10
SAMP:COUN 1000
CALC:SMO:RESP MED
CALC:SMO:STAT ON
CALC:STAT ON
READ?
Typical Response:
+9.9938382803,+9.9935231503,...
(1000 measurements)
l
Applies only to the 34465A and 34470A.
l
The
subsystem and the individual calculate functions are disabled for timestamp meas-
urements.
l
The instrument disables the smoothing (moving average) filter after a Factory Reset (
) or an Instru-
ment Preset (
).
Keysight True
volt
Series Operating and Service Guide
235
SCPI Programming Reference