DATA Subsystem
This subsystem allows you to configure and remove data from reading memory. The instrument clears all
measurements from reading memory when the measurement configuration changes, or when any of these
commands are executed:
,
,
Command Summary
DATA:LAST?
Returns the last measurement taken. You can execute this query at any time, even during a series of meas-
urements.
Parameter
Typical Return
(none)
(none)
Return the last measurement:
DATA:LAST?
Typical Response: One measurement with units. If no data is available, 9.91E37 (Not a Number) is returned
with units.
Example: +1.7300 VDC
DATA:POINts?
Returns the total number of measurements currently in reading memory. You can execute this query at
any time, even during a series of measurements.
Parameter
Typical Return
(none)
(none)
Return the number of measurements in reading memory:
DATA:POIN?
Typical Response: +215
l
You can store up to 1,000 measurements in the reading memory of the 34460A, 10,000 meas-
urements on the 34461A, 50,000 measurements on the 34465A/70A (
without
the MEM option), or
2,000,000 measurements on the 34465A/70A (
with
the MEM option).
SCPI Programming Reference
256
Keysight True
volt
Series Operating and Service Guide