[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe {<
range
>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe? [{MIN|MAX|DEF}]
Selects a fixed
voltage
range for frequency and period measurements.
This parameter is shared between frequency and period measurements. Setting or querying the parameter
with the FREQuency version of this command is identical to setting or querying it with the PERiod version.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the front and rear HI/LO
input terminals is 750 VAC (rms). The rms voltage is waveform dependent. A sine wave is limited to 750
VAC (rms), but a 1000 Vpk square wave is safe. Connection to AC mains power is further limited to CAT II
(300 V). See
Safety and Regulatory Information
for more information on the safety features and the
safe operation of this instrument.
Parameter
Typical Return
<
range
>: {100 mV|1 V|10 V|100 V|1000 V}. Default:
10V
+1.0403
Configures frequency measurements using the 10 VAC range. Make and read two measurements:
CONF:FREQ
FREQ:VOLT:RANG 10
SAMP:COUN 2
READ?
l
The input signal for frequency or period measurements has an
AC voltage
component. Use this com-
mand to select a fixed
voltage
range for frequency and period measurements. Use
{FREQuency|PERiod}:VOLTage:RANGe:AUTO
to disable or enable voltage autoranging or use
to select voltage autoranging by default.
l
Selecting a fixed range ([SENSe:]<
function
>:RANGe) disables autoranging.
l
If the input voltage is too large for the selected voltage range (manual ranging), the instrument dis-
plays the word Overload on the front panel and returns "9.9E37" from the remote interface. Autor-
anging can be enabled for the input voltage.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
342
Keysight True
volt
Series Operating and Service Guide