[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe] {ON|1|OFF|0}
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:POWer:LIMit[:STATe]?
Enables or disables low-power, 2-wire or 4-wire, RTD measurements.
This parameter is common to 2-wire and 4-wire RTD measurements. Using the FRTD version of this com-
mand or query is identical to using the RTD version.
Parameter
Typical Return
{ON|1|OFF|0}. Default: OFF.
0 (OFF) or 1 (ON)
Configure offset-compensated 2-wire RTD measurements. Make and read two measurements:
CONF:TEMP RTD
TEMP:TRANS:RTD:POW:LIM ON
TEMP:TRANS:RTD:OCOM ON
SAMP:COUN 2
READ?
Typical Response: +1.0501,+1.0701
l
Applies only to the 34465A and 34470A.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence] {<
reference
>|MIN|MAX|DEF}
[SENSe:]TEMPerature:TRANsducer:{FRTD|RTD}:RESistance[:REFerence]? [{MIN|MAX|DEF}]
Selects the nominal resistance (R0) for 2-wire and 4-wire RTD measurements. R0 is the nominal resistance
of an RTD at 0 °C.
Parameter
Typical Return
34460A/61A: 80 Ω to 120 Ω. Default: 100 Ω.
34465A/70A: 49 Ω to 2100 Ω. Default: 100 Ω.
+1.0002
Configure 4-wire RTD measurements using an RTD with an R
0
of 100.1 Ω. Make and read two meas-
urements:
CONF:TEMP FRTD
TEMP:TRAN:FRTD:RES 100.1
SAMP:COUN 2
READ?
Typical Response: +1.0402,+1.0402
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide
365
SCPI Programming Reference