[SENSe:]VOLTage:{AC|DC}:NULL:VALue {<
value
>|MIN|MAX|DEF}
[SENSe:]VOLTage:{AC|DC}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for voltage measurements.
This parameter setting is not shared between AC and DC measurements. The parameters are independent
for AC and DC measurements.
Parameter
Typical Return
-1200 to +1200 V. Default: 0.
+1.00000000E-02
Configure AC voltage measurements, using the null function to subtract 100 mV from the measurements.
Make and read two measurements:
CONF:VOLT:AC
VOLT:AC:NULL:STAT ON;VAL 100 mV
SAMP:COUN 2
READ?
Typical Response: +1.0400,+1.0400
l
Specifying a null value disables automatic null value selection (
).
l
To use the null value, the null state must be on (
[SENSe:]VOLTage:{AC|DC}:NULL:STATe ON
).
l
The null function is not available for DC ratio measurements.
l
This parameter is set to its default value after a Factory Reset (
), Instrument Preset
(
), or CONFigure function.