[SENSe:]VOLTage:{AC|DC}:RANGe {<
range
>|MIN|MAX|DEF}
[SENSe:]VOLTage:{AC|DC}:RANGe? [{MIN|MAX|DEF}]
Selects a fixed measurement range for AC and DC voltage measurements and DC ratio measurements.
This parameter setting is not shared between AC and DC measurements. The parameters are independent
for AC and DC measurements.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the front
and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform dependent.
A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is safe. Connection
to AC mains power is further limited to CAT II (300 V). See
for more information on the safety features and the safe operation of this
instrument.
Parameter
Typical Return
<
range
>: {100 mV|1 V|10 V|100 V|1000 V}
AC default: 10 V
DC default: 1000 V
+1.0001
Configure AC voltage measurements using the 100 V range. Make and read two measurements:
CONF:VOLT:AC
VOLT:AC:RANG 100
SAMP:COUN 2
READ?
Typical Response: +3.2401,+3.2401
l
Selecting a fixed range ([SENSe:]<
function
>:RANGe) disables autoranging.
l
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
plays the word
Overload
on front panel and returns "9.9E37" from the remote interface.
l
The instrument is set to the default range, with autoranging enabled (
), after a Factory Reset (
).
Keysight True
volt
Series Operating and Service Guide
377
SCPI Programming Reference