[SENSe:]VOLTage[:DC]:NPLC {<
PLC
>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for DC voltage and ratio measurements.
Integration time is the period that the instrument's analog-to-digital (A/D) converter samples the input sig-
nal for a measurement. A longer integration time gives better measurement resolution but slower meas-
urement speed.
Parameter
Typical Return
0.02, 0.2, 1, 10, 100. Default: 10 (34460A/61A)
0.02, 0.06, 0.2, 1, 10, 100. Default: 10 (34465A/70A without
DIG option)
0.001, 0.002, 0.006, 0.02, 0.06, .2, 1, 10, 100. Default: 10
(34465A/70A with DIG option)
See
for more information.
+1.0001
Configure DC voltage measurements using a 10 PLC integration time. Then make and read one meas-
urement:
CONF:VOLT:DC
VOLT:DC:NPLC 10
READ?
Typical Response: +6.27530000E-01
l
Only integration times of 1, 10, or 100 PLC provide normal mode (line frequency noise) rejection.
l
Setting the integration time also sets the measurement resolution.
shows the rela-
tionship between integration time and resolution. See
[SENSe:]VOLTage[:DC]:RESolution
.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
SCPI Programming Reference
382
Keysight True
volt
Series Operating and Service Guide